79 | ROLLING HILLS HEALTHCARE | 435035 | 2200 13TH AVE | BELLE FOURCHE | SD | 57717 | 2017-02-01 | 253 | D | 0 | 1 | X0TZ11 | Based on observation, interview, and manufacturer's instructions, the provider failed to have four of four different types of lifts on a preventative maintenance checklist. Findings include: 1. Random observations from 1/31/17 through 2/1/17 throughout the facility revealed: *There were four different brands and types of lifts in-use. -Arjo. -Invacare. -Medcare. -EZ lift. Interview on 2/1/17 at 1:00 p.m. with the maintenance supervisor and the administrator revealed they had not had any of the lifts on a preventative maintenance program. The certified nursing assistants would bring the lifts to the maintenance area if they were not working. They were unaware of what lifts needed clips and what lifts did not need them. Review of the manufacturer's instructions for the Arjo lifts revealed inspections should have been done monthly. Review of the manufacturer's instructions for the Invacare lifts revealed inspections should have been done monthly. Review of the manufacturer's instructions for the Medcare lifts revealed inspections should have been done daily and periodically. Review of the manufacturer's instructions for the EZ lifts revealed inspections should have been done every six months. | 2020-09-01 |